UV/VIS Spectrophotometer Ultrospec 4300 pro
Brand: Amersham Biosciences
Description: The Ultrospec 4300 pro is designed for high resolution scanning for assessment of impurities in sample preparations with bandwidth of less than 1.8 nm.
Specifications:
- Wavelength Range: 190-1100 nm, in 0.1nm steps
- Scanning Speed: 6200 nm/minute maximum, 1 nm step
- Monochromator: Concave grating with 1200 lines/mm
- Wavelength calibration: Automatic upon switch-on
- Spectral bandwidth: < 1.8nm
- Wavelength accuracy: ± 0.7nm
- Wavelength reproducibility: ± 0.2nm
- Light sources: Tungsten halogen and deuterium arc
- Detector: Single solid state silicon photodiode
- Photometric range: - 3.000 to + 3.000 A, 0.01 to 99999 concentration units, 0.1 to 200 %T
- Photometric linearity: ± 0.5% or ± 0.003 A to 3.000 A at 546nm, whichever is the greater
- Photometric reproducibility: Within 0.5% of absorbance value to 3.000A (at 546nm)
- Stray light typically: <0.025%T at 220nm using NaI, <0.025%T at 340nm using NaNO2
- Stability Warm Up: ± 0.001A/h at 340 nm at 0A after
- Noise: ± 0.001A near 0A and ± 0.002A near 2A at 546 nm
- Sample compartment size: 5.5" W x 8.6" D x 3.4" H(14 x 22 x 8 cm)